Environmental Monitoring System
RIGEL DATA Next
Modular EMS architecture engineered for GMP cleanrooms, particle, microbial, microclimate and vacuum, unified under one validated data layer.
- Modular. Seven hardware modules, configured not customized. Add capacity without rewriting architecture.
- Validated by design. GAMP 5, 21 CFR Part 11, EU GMP Annex 11 and ALCOA+ are embedded in the architecture, not bolted on.
- Open. Profinet downstream, OPC UA / Modbus / digital I/O upstream. No proprietary lock-in toward SCADA, BMS, MES or LIMS.
- Reduced footprint. Up to 50% lower in height vs typical rack-based EMS in equivalent configurations.
- One software platform. From a single isolator to a multi-department site, the same IEMS Next SCADA and configuration tool.
Contact us to learn more about the RIGELDATA Next
Environmental Monitoring System
Monitoring you can defend
A modular EMS architecture engineered for GMP cleanrooms — particle, microbial, microclimate and vacuum, unified under one validated data layer.
In a GMP cleanroom, the environmental monitoring system is the spine your validation rests on. Particle counters, microbial samplers, microclimate sensors and vacuum infrastructure must work together, and the data they produce has to be traceable, defensible in an audit, and integrated with the rest of the facility automation. Data Next is built to do exactly that, without forcing you to compromise on architecture, compliance or footprint.
How the system is organised
Seven hardware modules form the foundation of every Data Next installation. They are designed to be configured, not customised: facilities differ, but the building blocks remain the same.
Controller
Central PLC for data acquisition, alarm logic and SCADA communication.
Airflow
I/O
Vacuum Controller
Energy
PoE
Data Next OMNIA
Integrated Environmental Monitoring System
Architecture overview
At the centre of every Data Next installation is the Controller Module, which orchestrates the field modules over Profinet IRT. Each module is plug-and-play within the configured topology, and exposes its data upstream to the IEMS Next SCADA platform via OPC UA. Modbus TCP/RTU and digital I/O remain available for integration with BMS, MES, LIMS and Active Directory.
Small lab or compounding suite | 1× Data Next OMNIA | 1–4 |
Single cleanroom or isolator suite | 1× Controller + 2× Airflow + 1× Vacuum | 8–16 |
Mid-size facility, multi-area | 1× Controller + 6× Airflow + 2× I/O + 1× Vacuum + 1× PoE | 24–48 |
Multi-department site | Distributed Controllers + multiple Airflow / I/O / PoE + redundant Vacuum | 60+ |
Compliance is built in, not bolted on
GAMP 5 second edition, 21 CFR Part 11, EU GMP Annex 11 and ALCOA+. The audit trail is part of the system architecture, not a separate module. Every record is traceable to the user, the location and the conditions under which it was collected. The evidence an inspector is going to ask for is already there. Calibration of connected instruments is supported by Rigel’s ACCREDIA n. 01837, ISO/IEC 17025 accredited scope.
Footprint and deployment
Controller module
Central PLC for data acquisition, alarm logic and SCADA communication
Your monitoring system is only as defensible as the controller that runs it. The Data Next Controller Module is built around a Siemens CPU platform, the same architecture qualified across the pharmaceutical and life sciences industry. It coordinates the Airflow, I/O, Vacuum, PoE and Energy modules over Profinet and communication with the IEMS Next SCADA layer for real-time visualisation, alarm management and 21 CFR Part 11 compliant data integrity. Power is delivered by an internal 24 Vdc supply, with stackable or 19″ rack mounting. The architecture is open: Profinet for the field, OPC UA and Modbus toward the rest of your stack.
Features
- Powered by Siemens CPU platform
- Up to 5 configurable I/O sub-modules (AI / DI / DQ)
- IEMS Next / GE iFIX SCADA communication, 1-second data refresh
- 21 CFR Part 11 compliant audit trail at controller level
- Open architecture: Profinet, OPC UA, Modbus TCP/RTU
- Compact 2U Rigel enclosure
- Designed for cleanroom environments — EMC, RoHS, CE
Applications
- Centralised EMS for pharmaceutical cleanrooms
- Aseptic manufacturing and fill-finish facilities
- Multi-area / multi-department EMS deployments
- Integration into existing SCADA, BMS and MES stacks
- Migration from legacy EMS to modern architectures
- Hospital pharmacy compounding and ATMP suites
Technical specifications
Product code | Data Next Controller Module |
CPU | Siemens CPU |
Expansion | Up to 4 sub-modules (AI / DI / DQ in any combination) |
Communication | Profinet IRT downstream · OPC UA / Modbus TCP/RTU upstream |
Power supply | 100–240 Vac → 24 Vdc · 300 W output · internal industrial PSU |
Mains inlet | IEC C14 with fuse holder and rocker switch |
Mounting | Stackable or 2U — 19″ rack module |
Operating temperature | 0 to +55 °C · storage −30 to +60 °C, non-condensing |
Compliance | GAMP 5 · 21 CFR Part 11 · IEC 62368-1 · EMC · RoHS · CE |
Airflow Module
8-port manifold for particle and microbial sampling line management
Sampling reliability comes from the airflow path, not only from the counter. The Data Next Airflow Module is built around a Rigel-branded 8-channel custom manifold with built-in solenoid valves and an integrated mass flow meter on each line. Control is handled by a Siemens ET200SP head with analog inputs and digital outputs, so every valve actuation and every flow reading lives in the same Profinet data stream as the rest of the system.
Features
- 8-channel Rigel custom manifold, quick-connect fittings
- Internal 2-way solenoid valves
- Real-time mass flow meter, one per line
- Stackable design, VHP-compatible enclosure
- Internal 100–240 Vac → 24 Vdc / 150 W power supply
Applications
- Multi-point particle monitoring (multi-brand compatible)
- Active air microbial sampling with Hebas series (25 / 28.3 LPM)
- Mixed particle + microbial sampling lines under one manifold
- Aseptic fill-finish lines with continuous monitoring requirements
Technical specifications
Product code | Data Next Airflow Module |
Manifold | Rigel 8-port · 1/4″ G ports · plugs for unused ports |
Valves | 2-way solenoid · Ø7 mm · 24 Vdc |
Flow meter | Integrated mass flow meter · 1% accuracy |
Sampling flow rate | Up to 28.3 L/min per channel |
Power | 100–240 Vac → 24 Vdc · 300 W · IEC C14 inlet with fuse and switch |
Mounting | Stackable or 2U — 19″ rack module |
Operating temperature | 0 to +55 °C, non-condensing |
Compliance | GAMP 5 · IEC 62368-1 · CE · EMC · RoHS |
I/O Module
Scalable AI / DI / DQ expansion for sensors, alarms and actuators
Every facility wires environmental sensors differently. The Data Next I/O Module is built to absorb that variability without forcing custom hardware. Based on a Siemens ET200SP head, it covers 4–20 mA analog inputs, digital inputs and digital outputs in pre-defined configurations from 8 to 96 channels per module. Wiring is standardised on 16-pin push-in and feed-through connectors, so panel build and field commissioning move faster, and so service interventions do not depend on a single technician’s memory.
Features
- Mixed AI / DI / DQ configurations up to 96 channels
- Analog inputs 4–20 mA, isolated
- Digital inputs 24 Vdc
- Digital outputs 24 Vdc, 0.5 A
- Standardised 16-pin push-in connectors
- Hot-swappable Siemens ET200SP modules
Applications
- Microclimate sensor integration (T°, RH, ΔP)
- Door contact and pressure cascade monitoring
- Alarm beacon / horn / actuation
- Third-party sensor integration via 4–20 mA
- Distributed field cabinets requiring local I/O concentration
- Retrofit projects with mixed analog + digital signal inventory
- Dry contact exchange signal for third-part machine
Technical specifications
Product code | Data Next I/O Module |
Channels | 8 to 96 per module · AI / DI / DQ configurable |
Analog inputs | 4–20 mA, isolated, 16-bit resolution |
Digital inputs | 24 Vdc, sink/source configurable |
Digital outputs | 24 Vdc · 0.5 A per channel · short-circuit protected |
Power | 100–240 Vac → 24 Vdc · 300 W · IEC C14 inlet with fuse |
Mounting | Stackable or 2U — 19″ rack module |
Operating temperature | 0 to +55 °C, non-condensing |
Compliance | GAMP 5 · IEC 62368-1 · CE · EMC · RoHS |
Vacuum Controller Module
Centralised vacuum generation and diagnostics for sampling networks
Features
- Centralised vacuum pump control over Profinet
- Motor protection and soft-start management
- Continuous monitoring of motor status and line vacuum
- Interlock with Airflow Module sampling cycles
- Run-hour tracking for predictive maintenance
- Alarm routing to IEMS Next SCADA in real time
- Redundant pump configuration available on request
Applications
- Multi-room continuous particle monitoring networks
- Centralised vacuum systems serving up to 96 sampling points
- Aseptic processing facilities with 24/7 sampling
- Installations replacing dedicated pump-per-counter architectures
- Sites requiring redundant vacuum source for critical points
Technical specifications
Product code | Data Next Vacuum Controller Module |
Vacuum sensor | Huba Control 528 pressure transmitter (or equivalent) |
Vacuum range | 0 to −1 bar · configurable per application |
Pump compatibility | Becker, Busch, PWR, DVP and equivalent oil-free rotary vane pumps |
Diagnostics | Motor current · line vacuum · run hours · over-temperature |
Power | 24 Vdc internal · pump motor sized per project (typ. 400 Vac 3-phase) |
Mounting | Panel mount |
Compliance | GAMP 5 · IEC 62368-1 · CE · EMC · RoHS · UL 508A optional |
Energy Module
Cleanroom-grade power distribution with circuit protection and surge management
Features
- Centralised 100–240 Vac distribution to the module stack
- Per-line circuit protection with double fuse holder
- Single-point chassis grounding for clean signal reference
- Optional UPS interface for ride-through and graceful shutdown
- Mains entry via IEC C14 inlet with switch and integrated fuse
- Compatible with single- and dual-column rack configurations
Applications
- Centralised power for full Data Next stacks
- Sites requiring auditable circuit protection traceability
- Installations with site-level UPS backbone
- Multi-department deployments with distributed power architecture
- Facilities upgrading from distributed PSUs to centralised power management
Technical specifications
Product code | Data Next Energy Module |
Mains input | 100–240 Vac · 50/60 Hz · single-phase |
Output | 100-240 Vac |
Protection | Per-line fuse · thermal protection · surge protection |
Distribution | Up to 6 protected DC output lines |
UPS | Optional dry-contact / signal interface |
Mains inlet | IEC C14 with integrated fuse and rocker switch |
Mounting | Stackable or 2U — 19″ rack module |
Compliance | IEC 62368-1 · CE · EMC · RoHS |
PoE Module
Industrial Power over Ethernet for distributed sensor networks
Features
- 24 PoE++ ports (IEEE 802.3bt)
- Industrial-grade managed switch (VLAN, IGMP snooping, QoS)
- Centralised power for compatible remote sensors
- Backward compatibility with 802.3af / 802.3at devices
- Fibre uplinks available for long-run deployments
Applications
- Distributed networks of Aerix Remote⁺ particle counters
- New cleanroom builds aiming to reduce cabling footprint
- Segregated OT networks for environmental monitoring traffic
- Multi-room installations with shared switch infrastructure
Technical specifications
Product code | Data Next PoE Module |
Switch type | Industrial managed Layer-2 PoE++ |
PoE ports | 24 × RJ45 · IEEE 802.3bt |
Uplinks | Fibre / copper, configurable per project |
Power | 100–240 Vac internal supply |
Mounting | Stackable or 2U — 19″ rack module |
Diagnostics | Per-port link / PoE status · cable diagnostics · port mirroring |
Compliance | IEC 62368-1 · IEC 62443 (segmentation) · CE · EMC · RoHS |
Data Next OMNIA
Compact field node for reduced sampling installations and remote points
Features
- Compact wall-mount or table-top form factor
- Up to 4-port airflow manifold with valves and flow meters
- Profinet to centralised SCADA when present, standalone otherwise
- Integrated power protection
- Cleanroom-compatible enclosure
- Same SCADA and configuration tool as full Data Next
Applications
- Small or pilot cleanrooms (single area, 1–4 sampling points)
- Isolators, RABS, biosafety cabinets, single-point monitoring
- Satellite labs and remote QC areas
- Hospital compounding suites and operating rooms
- ATMP facilities with distributed small-volume areas
Technical specifications
Product code | Data Next OMNIA |
Airflow channels | AI 4–8 ch · DI 8–16 ch · DQ 8–16 ch (configurable) |
Communication | Profinet · Modbus TCP/RTU · OPC UA upstream |
Vacuum | External pump |
Power | 100–240 Vac, single-phase |
Mounting & enclosure | Wall-mount or table-top · 450 × 300 × 350 mm |
SCADA communication | Same IEMS Next platform as full Data Next |
Compliance | GAMP 5 · 21 CFR Part 11 · IEC 62368-1 · CE · EMC · RoHS |